Products Description
Dissolution test results can help pharmaceutical companies to optimise the dosage form design of a drug, such as choosing the right excipients and formulation process to ensure that the drug is released at the right rate in the body,In the field of pharmaceutical production and quality control, disintegrometer is a precision instrument, which is mainly used to detect the disintegration speed and disintegration of tablets, capsules and other solid preparations under the specified conditions. The use of disintegrator is of vital importance to ensure the quality of medicines and guarantee the safety of medication.
Features
Automation and intelligent control: The disintegrator adopts computer control technology with microcontroller as the core, and intelligently controls the water bath temperature, working time, lifting frequency and other performance parameters.
Precise temperature control: The disintegrator can automatically control the temperature of the water bath, which is usually set at 37.0℃ (Pharmacopoeia regulations), and the preset temperature can be reset at any time. The temperature control accuracy reaches ±0.5℃.
Independent basket system: The disintegrator is usually equipped with several sets of baskets, which can independently carry out disintegration experiments, such as the 4-basket working position system to determine the disintegration time limit.
Electronic Temperature Sensor: An electronic temperature sensor is provided to display and monitor the temperature at various points in the water bath and in the beaker to ensure the accuracy of the experiment.
Automatic lifting basket: The lifting frequency of the basket of the disintegrator is (30-31)/minute, and the lifting distance is (55±1)mm, which ensures that the samples are uniformly heated and stirred in the water bath.
Overheating alarm and automatic protection function: the instrument has a monitoring water bath temperature overheating alarm and automatic protection function to prevent the sample from being damaged by the high temperature.
Preset and Timing Function: The disintegrator allows the user to preset the timing range from 1 minute to 999 minutes with a time control accuracy of ±0.5 minutes.
Sieve design: The minimum distance from the sieve to the bottom of the cup is 25mm ± 2mm, and the aperture of the sieve is 2mm as standard, or 0.425mm or 1mm aperture can be selected.

Parameter
|
Name |
Disintegration tester |
|
The number of lifting baskets |
2 |
|
Number of glass tubes in hanging basket |
12 |
|
Volume of flat-bottomed beaker |
1000ml |
|
Basket shuttle frequency |
(30~32) times/minute |
|
Basket round trip travel |
(55±1) mm |
|
Minimum screen spacing |
(25±2) mm |
|
Timing range |
(1~900) points |
|
Timing accuracy |
±0.5min |
|
Temperature preset range |
(20~45)ºC |
|
Automatic temperature control accuracy |
±0.5°C |
|
Heater power |
500W |
|
Continuous working time |
More than 24 hours |
|
Power supply |
220V/50Hz |
|
Power |
600W |
|
Dimensions |
(40*30*42)cm |
application


| Disintegration equipment is used to test the disintegration speed and disintegration of tablets, capsules and other solid preparations under specified conditions, and is an important tool for pharmaceutical research and development, production process control and quality inspection. |
Hot Tags: tablet dissolution tester, China tablet dissolution tester manufacturers, suppliers, factory